Unusual pericuff leak around proseal-laryngeal mask airway corrected by clamping gastric drain port: A case report

Authors

  • Mohd. Mustahsin
  • Debesh Bhoi

DOI:

https://doi.org/10.32677/ijcr.v7i8.2973

Keywords:

Air leak, Laryngeal mask airway, Pericuff leak, ProSeal laryngeal mask airway

Abstract

ProSealTM-Laryngeal Mask Airway (PLMA) (Laryngeal Mask Company, Henley-on Thames, UK) is commonly used for securing the airway with an added advantage over classic LMA as its gastric drain tube allows the insertion of Ryle’s tube and suctioning of gastric contents. The ProSeal LMA is designed in such a way that it allows controlled mechanical ventilation. During controlled mechanical ventilation, air leaks can occur because of positive airway pressures. Air leaks from the gastric drain port are almost always due to the malposition of PLMA. Here, we report a case of air leak from gastric drain port despite correctly placed PLMA and its successful management without removing the device.

Downloads

Download data is not yet available.

Downloads

Published

2021-08-28

Issue

Section

Case Report

How to Cite

Unusual pericuff leak around proseal-laryngeal mask airway corrected by clamping gastric drain port: A case report. (2021). Indian Journal of Case Reports, 7(8), 345-347. https://doi.org/10.32677/ijcr.v7i8.2973

Most read articles by the same author(s)

<< < 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 > >>